Validation of Spectrometry Software: The Proactive Use of a Traceability Matrix in Spectrometry Software Validation, Part II: Practice - - Spectroscopy
FindAnalytichem Custom Search
About Search
 Home   Mass Spectrometry   ICP-MS   Infrared   FT-IR   UV-Vis   Raman   NMR   X-Ray   Fluorescence  
Make This Page Your Home Page!

Validation of Spectrometry Software: The Proactive Use of a Traceability Matrix in Spectrometry Software Validation, Part II: Practice


Spectroscopy



R.D. McDowall
In the first part of this column series on traceability matrices (1), we reviewed a system development life cycle model and then discussed the regulatory requirements and expectations as well as the business benefits of this document. We also looked at the terminology used and some of the principles of tracing requirements to other documents and activities in the life cycle.

Two points need to be reiterated from Part I for the discussion in this part. Requirements can either be tested or verified throughout the life cycle. Testable and verifiable were defined in Part I (1) as

Testable: The degree to which a requirement is stated in terms that permit establishment of test criteria and performance of tests to determine whether those criteria have been met. This is typically undertaken in the operational qualification (OQ) or performance qualification (PQ) phases of the life cycle.

Verifiable: The degree to which a requirement can be fulfilled by implementation (in the installation qualification phase), software configuration, writing standard operating procedures (SOPs), user training, vendor audit, calibration, or documentation.

This is important. In contrast to the GAMP 5 guide (2) that only discusses verification in the testing phases of a life cycle model, I prefer to split traceability into testing and verification because these are different, and in doing so, it makes the process easier and simpler to understand. When I talk about testing, I am referring to tasks that will be undertaken by the user (during the PQ) and also the vendor (during the OQ). In contrast, requirements that are verified are scattered throughout all the remaining stages of a life cycle and not just the testing ones.

Linking Requirements with Their Testing or Verification


Figure 1
To illustrate the principles of traceability between the requirements in the user requirements specifications (URS) and testing and verification in later stages of a system life cycle, let's look at Figure 1. The URS and the requirements that it contains are shown in the left-hand column. Across the top of the figure in a single row are the documents that could be written in a validation project. The figure caption lists what each of the abbreviations used in the table means. Before we start our traceability journey, the document names are what I would typically use; your organization might name them differently, but that is not a problem. This is to illustrate the principles of traceability in more detail than in Part I. You will need to interpret this approach using your terminology instead.

I will discuss the principles of traceability using most of the nine requirements listed in the URS in Figure 1. After the principles have been discussed, you can apply them to the remaining ones.

Requirement R1 is for the PC hardware upon which the spectrometer software will run and that the IT department will purchase. This is expanded in more detail to the make, model, memory, disk size, and configuration required to support the application and store the acquired data. It is traced first to item the specification T1 in a document called a technical specification; from here it is verified during the installation qualification (IQ) of the computer system. This answers the question: was what we specified actually delivered and installed, and if not, has the difference been documented.

R2 describes a requirement that needs an SOP written for the users to follow, and when this document is available the requirement will be verified. Collate all the user requirements that trace to procedures, check them against existing SOPs, and update as appropriate. Where no SOP exists, one must be written.

R3 leads to a function in the application software that will be configured and the setting is documented in the configuration specification in section 3.2 and tested during the PQ (for example, T). One can go further than just listing T under the PQ, but the detail to which we go will be discussed in a later section of this column.


Rate This Article
Your original vote has been tallied and is included in the ratings results.
View our top pages
Average rating for this page is: 0
Post a Comment
Your email address will NOT be published.
appears with your comment
read our privacy policy
Note: does not support HTML
All comments submitted are subject to review, and may be delayed before posting. We reserve the right not to post comments.
Headlines from LCGC North America and Chromatography Online
German office
ChromaTOF 4.32
analytica China 2010
Compilation of clinical applications
Developing Malaysian biotechnology
Source: Spectroscopy,
Click here